![ILX Lightwave FPM-8200 Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/ilx-lightwave/fpm-8200/fpm-8200_user-manual_2044775058.webp)
G P I B C O M M A N D R E F E R E N C E
GPIB Command Reference Summary
42
FPM-8200
C H A P T E R
4
ERRors?
Errors?
Request errors since last ERR? request
*ESE <nrf value>
Set standard event status
enable register
Set the standard event status enable register
*ESE?
Standard event status
enable register?
Request standard event status enable register
*ESR?
Standard event
status register?
Request the standard event status register
EVEnt?
Event status register?
Requests the event status register
FILTer FAST
Fast filter
Select fast averaging and display update
FILTer MED
Medium filter
Select medium averaging and display update
FILTer SLOW
Slow filter
Select slow averaging and display update
FILTer?
Filter?
Request the measurement filter setting
*IDN?
Identify?
Request the FPM−8200 to identify itself
MESsage <ASCII
string>
Message
Store character string into message memory
MESsage?
Message?
Request character string from message memory
MODE?
Mode?
Request the measurement mode
MODE:DB
dB mode
Select decibels (dB) relative measurement mode
MODE:DBM
dBm mode
Select logarithmic (dBm) measurement mode
MODE:W
Watts mode
Select linear (watts) measurement mode
*OPC
Operation complete
Set operation complete flag
*OPC?
Operation complete?
Request operation complete status
POWer?
Power?
Request the value of measured optical power
*PSC <nrf value>
Power−on status clear
Set automatic power−on register clearing
*PSC?
Power−on status clear?
Requests status of power−on status clear flag
*PUD?
Protected user data?
Request the factory−stored identification string
RADix BIN
Binary radix
Set binary numeric responses
RADix DEC
Decimal radix
Set decimal numeric responses (default)
RADix HEX
Hex radix
Set hexadecimal numeric responses
RADix OCT
Octal radix
Set octal numeric responses
RADix?
Radix?
Request the radix of numeric responses
RANge <nrf value>
Range
Set the photodetector current gain range
RANge?
Range?
Request the photodetector current gain range
RANge:AUTO <nrf
value>
Autorange on/off
Set AUTO or MANUAL ranging mode
RANge:AUTO?
Autorange?
Request status: AUTO or MANUAL range
*RCL <nrf value>
Recall setup
Recall a stored setup configuration
REF <nrf value>
Set reference
Set a reference level in dBm
REF?
Reference value?
Request the reference level value
RESP?
Responsivity?
Request the calibrated detector responsivity data
*RST
Reset
Recall setup #0, set OCIS and OQIS
Table 4.2
GPIB Summary Reference Command List
Syntax
Meaning
Function
Содержание FPM-8200
Страница 2: ......
Страница 8: ...L I S T O F FIG U R E S vi FPM 8200...
Страница 10: ...L I S T O F TA B L E S viii FPM 8200...
Страница 16: ...WA R R A N T Y xiv FPM 8200...
Страница 36: ...FR O NT PA N E L O P E R A T I O N Operating Guidelines 20 FPM 8200 C H A P T E R 2...
Страница 90: ...TR O U B L E S H OO T I N G A N D M A I N T E N A N C E Error Messages 74 FPM 8200 C H A P T E R 5...
Страница 96: ...80 FPM 8200 A P P E N D I X A...