Linear Power Mode Display
Depending on the input power level, the FPM-8200 in linear mode displays power
in units nW, µW, or mW. Within these ranges, the decimal point is moved as
appropriate for the value:
To keep the display from rapidly changing when measuring at one of these
boundaries, the display in linear mode includes 1% hysteresis. To illustrate, here
is an example of changing from 99 nW to 100 nW (1% = 1 nW):
Power (watts)
<100 x 10
−9
100 x 10
−9
to
<1 x 10
−6
1 x 10
−6
to
<100 x 10
−6
100 x 10
−6
to
<1 x 10
−3
1 x 10
−3
to
<100 x 10
−3
100 x 10
−3
to
<1W
1W
Units
nW
nW
μW
μW
mW
mW
mW
Decimal places
3
2
3
2
3
2
1
Example
85.436 nW
327.24 nW
38.361 μW
542.76 μW
38.974 mW
756.58 mW
1062.3 mW
Increasing
Decreasing
98.999
100.01
99.000
100.00
99.001
99.99
99.002
99.98
.
99.97
.
.
99.998
.
99.999
99.01
100.00
99.00
100.01
98.999
100.02
98.998
Содержание FPM-8200
Страница 2: ......
Страница 8: ...L I S T O F FIG U R E S vi FPM 8200...
Страница 10: ...L I S T O F TA B L E S viii FPM 8200...
Страница 16: ...WA R R A N T Y xiv FPM 8200...
Страница 36: ...FR O NT PA N E L O P E R A T I O N Operating Guidelines 20 FPM 8200 C H A P T E R 2...
Страница 90: ...TR O U B L E S H OO T I N G A N D M A I N T E N A N C E Error Messages 74 FPM 8200 C H A P T E R 5...
Страница 96: ...80 FPM 8200 A P P E N D I X A...