![ILX Lightwave FPM-8200 Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/ilx-lightwave/fpm-8200/fpm-8200_user-manual_2044775033.webp)
FR O N T PA N E L O P E R A T I O N
Operating Guidelines
01_10
FPM-8200
17
C H A P T E R
2
Operating Guidelines
The discussion below presents guidelines for operation, as well as some common
operating procedures. Remote operations are discussed in the next chapter.
Warm-up and Environmental Considerations
To achieve rated accuracy, let the FPM-8200 warm up for at least 1 hour before
use. Operate the meter within the environmental limits specified in Chapter 1. Best
accuracy will be available for operation near the calibration temperatures.
Summary of Operating Procedures
The following list is a summary of procedures discussed in more depth earlier in
this chapter.
1
Cover the input connector with the supplied cap and press
ZERO
.
2
Install the appropriate fiber adaptor on the front panel
INPUT
connector. Connect your
input fiber patchcord.
3
Recall your setup parameters by pressing
RECALL
, or by setting the parameters
individually as described in the following steps.
4
To set measurement parameters:
4a
Press
and adjust the wavelength by pressing
UP
or
DOWN
.
4b
Press
AUTO/MAN
to select either auto-ranging or manual gain range. In manual gain
mode, pressing UP or DOWN steps the gain by a factor of 10. To set a particular gain
range, see Gain Ranges on page 7.
In Manual Range mode, display "
- OL -
" indicates the input power is greater than 99.1% of
range. In Auto Range mode, this indicates input power is greater than
1.59 dBm.
In Manual Range mode, display "
- LO -
" indicates the input power is less than 5% of
range. In Auto Range mode, this indicates input power is less than -76 dBm.
4c
Press
dB
and
LOCAL/ADR
together to check the USER CAL gain factor. Normally
this factor should be 1.000. You can change the gain factor while it is displayed by
pressing
UP
or
DOWN
.
5
To save a new setup, press
SAVE
followed by
UP
or
DOWN
for a memory number (1
through 10), then press
SAVE
again. Whether or not you save the setup, at next power-up
the FPM-8200 will return to the same setup as you left it.
Measuring Higher Power
The FPM-8200 will not be damaged by input power up to 10 dBm. However,
linearity generally moves outside specification limits above the upper specified
input power limit (see Table 1.1 on page 4).
Содержание FPM-8200
Страница 2: ......
Страница 8: ...L I S T O F FIG U R E S vi FPM 8200...
Страница 10: ...L I S T O F TA B L E S viii FPM 8200...
Страница 16: ...WA R R A N T Y xiv FPM 8200...
Страница 36: ...FR O NT PA N E L O P E R A T I O N Operating Guidelines 20 FPM 8200 C H A P T E R 2...
Страница 90: ...TR O U B L E S H OO T I N G A N D M A I N T E N A N C E Error Messages 74 FPM 8200 C H A P T E R 5...
Страница 96: ...80 FPM 8200 A P P E N D I X A...