![ILX Lightwave FPM-8200 Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/ilx-lightwave/fpm-8200/fpm-8200_user-manual_2044775048.webp)
R E M O T E O P E R A T I O N
Internal Registers
32
FPM-8200
C H A P T E R
3
Register Structure and Contents
This section provides information about the structure and contents of the different
internal registers.
Standard Event Status Register Pair
Standard Event Status is an 8-bit register/enable-register pair with this structure:
Figure 3.5
Standard Event Status
The Standard Event Status Enable Register allows you to control which events in
the Standard Event Status Register can set bit 5 in the status byte register. The
format of this register pair is defined by GPIB standards.
Standard event status indicates the FIRST occurrence, and does not change
except when reset. The FPM-8200 starts at power ON with the standard event
status register set to all zeros. Querying the Standard Event Status Register
resets its contents to zeros.
Bit 0 is key to an interrupt-driven control program, because it allows you to
generate a service request (SRQ) when current operations are completed before
making a measurement. To generate an operation complete SRQ, follow this
sequence:
•
*ESE 1
- Enable bit 0 of the Standard Event Status Enable Register.
•
*SRE 32
- Enable bit 5 of the Service Request Enable Register.
•
*OPC
- Set bit 0 in the Event Status Register when all pending overlapped commands have
been completed. (See
*OPC
in Chapter Four).
Operation complete means:
•
The FPM-8200 internal controller is idle.
•
A measurement result is available.
7
6
5
4
3
2
1
0
7
6
5
4
3
2
1
0
AND
AND
AND
AND
AND
AND
AND
OR
Standard Event Status Register
Standard Event Status Enable Register
To Bit 5,
Status Byte
Register
Operation Complete
Query Error
Device Error
Execution Error
Command Error
User Request
Power ON
Not used
*ESR?
*ESE <nrf>
*ESE?
Содержание FPM-8200
Страница 2: ......
Страница 8: ...L I S T O F FIG U R E S vi FPM 8200...
Страница 10: ...L I S T O F TA B L E S viii FPM 8200...
Страница 16: ...WA R R A N T Y xiv FPM 8200...
Страница 36: ...FR O NT PA N E L O P E R A T I O N Operating Guidelines 20 FPM 8200 C H A P T E R 2...
Страница 90: ...TR O U B L E S H OO T I N G A N D M A I N T E N A N C E Error Messages 74 FPM 8200 C H A P T E R 5...
Страница 96: ...80 FPM 8200 A P P E N D I X A...