FR O N T PA N E L O P E R A T I O N
The FPM-8200 Front Panel
8
FPM-8200
C H A P T E R
2
Wavelength
Because the detector responds differently depending on wavelength, it is
important to set the correct wavelength.
To set the wavelength:
1
Press
to enable the wavelength settin, illuminating the SET indicator below the
wavelength display.
2
While the SET indicator is on, press
UP
or
DOWN
to change the wavelength setting.
For more information concerning wavelength, see Photodetector Responsivity on
page 19.
Save and Recall
Use SAVE and RECALL to store and reuse FPM-8200 setup parameters for future
use. These parameters include the gain range, wavelength, display mode, filter
mode, and user calibration offset value. Stored setups save time and reduce the
chance of setup error for tests that are repeated periodically.
Setup parameters are stored in non-volatile memory, so they are always available
when you turn-on the instrument. You can save your own setup parameters in any
of ten memory locations numbered from 1-10. When you RECALL that number,
the FPM-8200 reconfigures to the stored setup parameters.
In addition, the FPM-8200 automatically stores two other configurations. When
you turn off the FPM-8200, the electrical current settings are stored and
automatically recalled the next time you turn-on the FPM-8200. This happens
whether or not you save the setup. Additionally, there is a consistent default
configuration, stored as "0" (zero). Setup 0 sets the following configuration:
Note:
Note that the reference value is stored in volatile memory. Therefore, the reference
is returned to 0.000 dBm upon any recall operation or a power down/up sequence.
If you recall setup zero by the front panel, GPIB mode is set to LOCAL. If you use
GPIB to recall setup 0 (
*RCL 0
), GPIB mode is set to REMOTE.
Function
Setting
GPIB Equivalent
Display mode
Linear (nW, μW, mW)
MODE: W
Wavelength
1550 nm
WAVE 1550
Filter mode
Medium
(average 10 measurements)
FILTER MED
Range mode
AUTO
RANGE:AUTO
User cal value
1.000
(indicator off)
CAL:USER 1
Содержание FPM-8200
Страница 2: ......
Страница 8: ...L I S T O F FIG U R E S vi FPM 8200...
Страница 10: ...L I S T O F TA B L E S viii FPM 8200...
Страница 16: ...WA R R A N T Y xiv FPM 8200...
Страница 36: ...FR O NT PA N E L O P E R A T I O N Operating Guidelines 20 FPM 8200 C H A P T E R 2...
Страница 90: ...TR O U B L E S H OO T I N G A N D M A I N T E N A N C E Error Messages 74 FPM 8200 C H A P T E R 5...
Страница 96: ...80 FPM 8200 A P P E N D I X A...