Intended use

Use:

The

 OVEN 125 basic - dry

 / 

OVEN 125 basic - dry glass

 

is a drying oven for a wide range of heating, drying and heat-
ing application�
The device may only be used for heating up non-explosive 
and non-flammable substances and objects� Do not process 
substances or objects that might cause explosive vapor�

Range of use:

Indoor environments similar to that a laboratory of research, 
teaching, trade or industry�

The safety of you cannot be guaranteed:
-  If the device is operated with accessories that are not sup-

plied or recommended by 

IKA

-  If the device is operated improperly or in contrary to the 

IKA

 specifications�

-  If the device or the printed circuit board are modified by 

third parties�

Useful information

Calibration and adjustment:

This menu shows how offset values can be entered after the respective temperatures are approached and with an external 
measuring device cross-checked (calibrated)�

[OC]

300

250

200

150

100

50

0

3-point calibration and offset

factory calibration

CAL 1

-5K

CAL 3

+5K

CAL 2

+7K

resulting offset graph

offset < 10 K
deviation between offset points > 30 K

Individual load and ambient condition might influent the temperature behaviour of the chamber� To compensate that influ-
ence there’s the option to neutralize by using the 3-point calibration and offset function� This function allows the tempera-
ture to be calibrated and adjusted at three different temperature values� 

Use a calibrated reference measuring device in case of adjustment�
The operation can be done at a single, or up to three temperature levels: 

CAL 1:

 Temperature calibration and adjustment at low value of temperature� 

CAL 2:

 Temperature calibration and adjustment at medium value of temperature� 

CAL 3:

 Temperature calibration and adjustment at high value of temperature�

Temperature control detection:

During operating, accidental door openings could cause an unexpected temperature drop and activate the “door” 
message and also an alarm tone sounds�

Switch off the display message and stop the alarm sound by pressing any key�

Note:

 The alarm and key tone can be deactivated at the menu settings�

Содержание Oven 125 basic dry

Страница 1: ...Operating instructions EN OVEN 125 basic dry OVEN 125 basic dry glass...

Страница 2: ...ion 1 Mains switch 2 Door handle 3 Device foot 4 Operator Panel and display 5 USB interface and temperature limit 6 Mains socket 7 Venting flap Device setup Dangerous parts Dangerous parts Device setu...

Страница 3: ...bols 4 Safety instructions 4 Intended use 6 Useful information 6 Transportation and storage 7 Unpacking 7 Installation 7 Operator panel and display 11 Setting the temperature limit 12 Commissioning 12...

Страница 4: ...an be accessed by everyone Ensure that only trained staff work with the device Follow the safety instructions guidelines occupational health and safety and accident prevention regulations Transport an...

Страница 5: ...The voltage stated on the type plate must correspond to the mains voltage The device can only be disconnected by pulling out the mains plug Socket must be earthed The device will automatically restar...

Страница 6: ...libration and offset factory calibration CAL 1 5K CAL 3 5K CAL 2 7K resulting offset graph offset 10 K deviation between offset points 30 K Individual load and ambient condition might influent the tem...

Страница 7: ...and corrosive ambient air during transportation and storage It is also important to ensure that the relative humidity does not exceed 80 The device must be completely emptied before transporta tion an...

Страница 8: ...l The screws used for fixing is selected according to the wall conditions and is not included As indicated in following figures attach the device to a wall with the tilt protections included with the...

Страница 9: ...device Never stack more than two devices When stacking the device four persons are required to lift it In order to keep enough space please use stacking adapter recommend by IKA 150 mm 150 mm Stacking...

Страница 10: ...avy about 15 kg of the door Open the door Hold the door and loosen the hinge pin with a suitable tool Remove the door carefully Alternate the position of the keepers and hinges with 4 mm hexagonal wre...

Страница 11: ...Timer Indicate the timer function is activated K Timer key Increase the timer setting value Navigation selecting the settings in the menu L Timer key Decrease the timer setting value Navigation selec...

Страница 12: ...function is activated by pressing the Start Stop key I the heating indicator T lights up Then the target temperature and the actual temperature appears on display A alternately Accordingly setting val...

Страница 13: ...hts up By pressing the lock key G about 2 seconds again the con trol elements are released LED F no longer lights up Switching between Counter and Timer function The Counter function could be switch t...

Страница 14: ...result This function allows the temperature to be calibrated and adjusted to three dif ferent temperature values When calibration locate a calibrated reference measuring sensor at the center of the c...

Страница 15: ...t side of the screen and actual temperature appears on the left side of the screen Wait about 1 2 hours until temperature is stable e g reach about 98 8 C and Set LED S no longer flashes and the follo...

Страница 16: ...ndations for the design of electrical plug connections for analogue and digital signal transmission on individual items of laboratory control equipment rev 1 1 The NAMUR commands and the additional sp...

Страница 17: ...Maintenance and cleaning The device is maintenance free It is only subject to the natural wear and tear of components and their statistical failure rate Cleaning For cleaning disconnect the mains plug...

Страница 18: ...No internal sensor Heating off Contact IKA service department If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps Contact the se...

Страница 19: ...display LED Max timer counter 100 days d 99 23 59 hh mm Max number of tray 6 Max load for one tray kg 30 Interface USB Permissible ambient temperature C 5 40 Permitted ambient humidity 80 Permitted o...

Страница 20: ...s lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ika cn POLAND IKA Poland Sp z o o Phone 48 22 201 99 79 eMail sales poland ika com USA IKA Works Inc Phone 1 910 452 7059 eMail...

Отзывы: