Operating mode:
Mode A:
The device doesn’t restart again after power on/
power failure� The setting of heating temperature is adjustable�
Mode B:
The device restart again after power on/power
failure� The setting of heating temperature is adjustable�
Mode C:
The device restart again after power on/power
failure� The setting of heating temperature is not adjustable�
Alarm and key tone:
Beep on:
alarm and key tone is activated�
Beep off:
no alarm and key tone�
Factory reset:
No reset�
Reset�
Calibration and adjustment:
The device is calibrated and adjusted in the factory�
Individual sample loading and ambient condition might in-
fluent the temperature behaviour of the chamber� You can
select and make 1-point, 2-point or 3-point calibration and
adjustment as necessary�
By using 3-point calibration to compensate that influence,
the device can get the optimal result� This function allows
the temperature to be calibrated and adjusted to three dif-
ferent temperature values�
When calibration, locate a calibrated reference measuring
sensor at the center of the cabinet�
Calibration information:
In this menu option, you can review the calibration informa-
tion in the device:
Enter the calibration information by pressing “Start/Stop”
key (I) and following screens appear:
If the device is selected 3-point calibration, following cali-
bration information screen will appear:
Back to calibration information screen by pressing “Start/
Stop” key (I)�
By pressing “Timer (+)” key (K) or “Timer (-)” key (L), you can
switch between following three screens and review relate
calibration information:
Edit calibration parameters:
In this menu option, you can edit the calibration parameter
in the device:
Confirm to edit the setting values by pressing “Start/Stop”
key (I) and following screen appear:
By pressing “Temp (+)” key (O) or “Temp (-)” key (P), you can
select following types calibration:
(no adjustment)
(1-point calibration)
(2-point calibration)
(3-point calibration)