background image

5

CAUTION

 

There is a crushing risk when open

ing or closing the door� 

Keep your hands away from squeezing edges�

 

Set up the device in a spacious area on an even, stable, 

clean, non-slip, dry and fireproof surface�

 

Secure the device with the anti-tilt device to ensure stabil-

ity while loading and unloading shelves and during opera-
tion� Ensure the anti-tilt bracket is properly fixed, in par-
ticular when the device is stacked�

 

The feet of the equipment must be clean and undamaged�

Work with the device:

WARNING

• Do not start up the device if:

- it is damaged
- cable is damaged�

• The device must have cooled down before loading/un-

loading� 

• Do not move any loaded racks�
• Heavy loads can lead to tilting of the device� Observe 

loading recommendations for the shelves�

• Never use the device for:

- heating food
- drying living beings
- heating room

• The device has a temperature limiter (TB) class 1 acc� 

DIN 12880� This means that in case of a single failure 
the device is switched off at 270°C� The temperature 
may be higher for a short time in the area of the heater 
and on the walls� So beware of risk of fire or explosion 
of unsuitable load�

CAUTION

• The area around the air outlet nozzle can get very hot at 

high operating temperatures�

Power supply/Switching off the device:

• The voltage stated on the type plate must correspond to 

the mains voltage�

• The device can only be disconnected by pulling out the 

mains plug�

• Socket must be earthed�
• The device will automatically restart in mode B and C fol-

lowing any interruption to the power supply�

• Protect the power cable against kinks�
• Check power cables at regular intervals for possible mate-

rial fatigue�

• Power cable should not come into contact with hot parts�

Heating:

CAUTION

• When high operating temperatures are used, the device 

may get very hot, particularly at the openings at the back�

• Risk of burns caused by vapour or hot surfaces� Pay atten-

tion to the residual heat after switching off�

Accessories:

 NOTICE

• Use the locking brake when using the roller frame or roller 

stand�

• Safe operation is only guaranteed with the accessories de-

scribed in the “Accessories” section

• Only use IKA approved accessories!
• Use only original IKA spare parts!
• Check the device and accessories for damage before each 

use� Do not use damaged components�

• Protect the device and accessories from blows and im-

pacts�

Maintenance:

• The device must only be opened by trained specialists, 

even during repair� The device must be unplugged from 
the power supply before opening� Live parts inside the de-
vice may still be live for some time after unplugging from 
the power supply�

Dis

posal instructions:

• The device, accessories and packaging must be disposed of 

in accordance with local and national regulations�

Содержание Oven 125 basic dry

Страница 1: ...Operating instructions EN OVEN 125 basic dry OVEN 125 basic dry glass...

Страница 2: ...ion 1 Mains switch 2 Door handle 3 Device foot 4 Operator Panel and display 5 USB interface and temperature limit 6 Mains socket 7 Venting flap Device setup Dangerous parts Dangerous parts Device setu...

Страница 3: ...bols 4 Safety instructions 4 Intended use 6 Useful information 6 Transportation and storage 7 Unpacking 7 Installation 7 Operator panel and display 11 Setting the temperature limit 12 Commissioning 12...

Страница 4: ...an be accessed by everyone Ensure that only trained staff work with the device Follow the safety instructions guidelines occupational health and safety and accident prevention regulations Transport an...

Страница 5: ...The voltage stated on the type plate must correspond to the mains voltage The device can only be disconnected by pulling out the mains plug Socket must be earthed The device will automatically restar...

Страница 6: ...libration and offset factory calibration CAL 1 5K CAL 3 5K CAL 2 7K resulting offset graph offset 10 K deviation between offset points 30 K Individual load and ambient condition might influent the tem...

Страница 7: ...and corrosive ambient air during transportation and storage It is also important to ensure that the relative humidity does not exceed 80 The device must be completely emptied before transporta tion an...

Страница 8: ...l The screws used for fixing is selected according to the wall conditions and is not included As indicated in following figures attach the device to a wall with the tilt protections included with the...

Страница 9: ...device Never stack more than two devices When stacking the device four persons are required to lift it In order to keep enough space please use stacking adapter recommend by IKA 150 mm 150 mm Stacking...

Страница 10: ...avy about 15 kg of the door Open the door Hold the door and loosen the hinge pin with a suitable tool Remove the door carefully Alternate the position of the keepers and hinges with 4 mm hexagonal wre...

Страница 11: ...Timer Indicate the timer function is activated K Timer key Increase the timer setting value Navigation selecting the settings in the menu L Timer key Decrease the timer setting value Navigation selec...

Страница 12: ...function is activated by pressing the Start Stop key I the heating indicator T lights up Then the target temperature and the actual temperature appears on display A alternately Accordingly setting val...

Страница 13: ...hts up By pressing the lock key G about 2 seconds again the con trol elements are released LED F no longer lights up Switching between Counter and Timer function The Counter function could be switch t...

Страница 14: ...result This function allows the temperature to be calibrated and adjusted to three dif ferent temperature values When calibration locate a calibrated reference measuring sensor at the center of the c...

Страница 15: ...t side of the screen and actual temperature appears on the left side of the screen Wait about 1 2 hours until temperature is stable e g reach about 98 8 C and Set LED S no longer flashes and the follo...

Страница 16: ...ndations for the design of electrical plug connections for analogue and digital signal transmission on individual items of laboratory control equipment rev 1 1 The NAMUR commands and the additional sp...

Страница 17: ...Maintenance and cleaning The device is maintenance free It is only subject to the natural wear and tear of components and their statistical failure rate Cleaning For cleaning disconnect the mains plug...

Страница 18: ...No internal sensor Heating off Contact IKA service department If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps Contact the se...

Страница 19: ...display LED Max timer counter 100 days d 99 23 59 hh mm Max number of tray 6 Max load for one tray kg 30 Interface USB Permissible ambient temperature C 5 40 Permitted ambient humidity 80 Permitted o...