Setting up:
Fixation against the wall is compulsory!
- Place the device on and even, clean, dry and fireproof surface in well-ventilated environment�
- Keep sufficient distance to radiators and other heat sources�
- Keep at least 150 mm between walls and the device�
Fixing the device to the wall:
The screws used for fixing is selected according to the wall conditions and is not included�
As indicated in following figures, attach the device to a wall with the tilt protections included with the device�
Fasten the tilt protections to the rear side of the device�
Inserting the shelves:
- Open the door�
- Insert the shelves�
- Load shelves evenly�
B
WARNING
NOTICE
Note:
Keep distance between
samples and hot inner surface�
Temperature of the inner wall
may be much higher than set
temperature!