“Watchdog” function, monitoring the serial data flow:
The following applies to situations where the watchdog function is enabled (see Namur instructions)� If no new transmissions 
of these commands from the PC take place within the preset watchdog time, the heating function will be shutdown 
according to the watchdog mode selected or will be controlled using the preset setpoints� An operating system crash, a PC 
power failure or a fault in the connecting cable to the device can cause an interruption in data transmission�

“Watchdog”– Mode 1
If an interruption in data transmission occurs which is longer than the preset watchdog time, the heating function will be 
shutdown and the error message PC 1 will be displayed�

“Watchdog”– Mode 2
If an interruption in data transmission occurs which is longer than the preset watchdog time, the temperature setpoint will 
be set to the WD safety temperature setpoint� The PC 2 warning message will be displayed�

Connections between device and external devices:

USB 2�0 cable (A - B):
This cable is required to connect USB interface to a PC�

Maintenance and cleaning

The device is maintenance-free� It is only subject to the natural 
wear and tear of components and their statistical failure rate�

Cleaning:

For cleaning disconnect the mains 
plug!

Use only cleaning agents which have been approved by 

IKA

 

to clean 

IKA

 devices: Water containing surfactant / isopropyl 

alcohol�

-  Wear protective gloves while cleaning the device�
- Electrical devices may not be placed in the cleansing agent 

for the purpose of cleaning�

-  Do not allow moisture to get into the device when cleaning�
-  If a different cleaning or decontamination method than 

the method defined by 

IKA

 is planned, you must ascertain 

with 

IKA

 that this method does not damage the device�

Spare parts order:

When ordering spare parts, please give:
-  Machine type�
-  Serial number, see type plate�
-  Item and designation of the spare part, 
 see 

www.ika.com

, spare parts diagram and list�

Repair:

Please send the device for repair only after it has 

been cleaned and is free from any materials which 

may constitute a health hazard.

For repair, please request the “Decontamination Certificate” 
from  

IKA

, or download printout of it from the 

IKA

 website 

www.ika.com

If you require servicing, return the device in its original pack-
aging� Storage packaging is not sufficient� Please also use 
suitable transport packaging�

Содержание Oven 125 basic dry

Страница 1: ...Operating instructions EN OVEN 125 basic dry OVEN 125 basic dry glass...

Страница 2: ...ion 1 Mains switch 2 Door handle 3 Device foot 4 Operator Panel and display 5 USB interface and temperature limit 6 Mains socket 7 Venting flap Device setup Dangerous parts Dangerous parts Device setu...

Страница 3: ...bols 4 Safety instructions 4 Intended use 6 Useful information 6 Transportation and storage 7 Unpacking 7 Installation 7 Operator panel and display 11 Setting the temperature limit 12 Commissioning 12...

Страница 4: ...an be accessed by everyone Ensure that only trained staff work with the device Follow the safety instructions guidelines occupational health and safety and accident prevention regulations Transport an...

Страница 5: ...The voltage stated on the type plate must correspond to the mains voltage The device can only be disconnected by pulling out the mains plug Socket must be earthed The device will automatically restar...

Страница 6: ...libration and offset factory calibration CAL 1 5K CAL 3 5K CAL 2 7K resulting offset graph offset 10 K deviation between offset points 30 K Individual load and ambient condition might influent the tem...

Страница 7: ...and corrosive ambient air during transportation and storage It is also important to ensure that the relative humidity does not exceed 80 The device must be completely emptied before transporta tion an...

Страница 8: ...l The screws used for fixing is selected according to the wall conditions and is not included As indicated in following figures attach the device to a wall with the tilt protections included with the...

Страница 9: ...device Never stack more than two devices When stacking the device four persons are required to lift it In order to keep enough space please use stacking adapter recommend by IKA 150 mm 150 mm Stacking...

Страница 10: ...avy about 15 kg of the door Open the door Hold the door and loosen the hinge pin with a suitable tool Remove the door carefully Alternate the position of the keepers and hinges with 4 mm hexagonal wre...

Страница 11: ...Timer Indicate the timer function is activated K Timer key Increase the timer setting value Navigation selecting the settings in the menu L Timer key Decrease the timer setting value Navigation selec...

Страница 12: ...function is activated by pressing the Start Stop key I the heating indicator T lights up Then the target temperature and the actual temperature appears on display A alternately Accordingly setting val...

Страница 13: ...hts up By pressing the lock key G about 2 seconds again the con trol elements are released LED F no longer lights up Switching between Counter and Timer function The Counter function could be switch t...

Страница 14: ...result This function allows the temperature to be calibrated and adjusted to three dif ferent temperature values When calibration locate a calibrated reference measuring sensor at the center of the c...

Страница 15: ...t side of the screen and actual temperature appears on the left side of the screen Wait about 1 2 hours until temperature is stable e g reach about 98 8 C and Set LED S no longer flashes and the follo...

Страница 16: ...ndations for the design of electrical plug connections for analogue and digital signal transmission on individual items of laboratory control equipment rev 1 1 The NAMUR commands and the additional sp...

Страница 17: ...Maintenance and cleaning The device is maintenance free It is only subject to the natural wear and tear of components and their statistical failure rate Cleaning For cleaning disconnect the mains plug...

Страница 18: ...No internal sensor Heating off Contact IKA service department If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps Contact the se...

Страница 19: ...display LED Max timer counter 100 days d 99 23 59 hh mm Max number of tray 6 Max load for one tray kg 30 Interface USB Permissible ambient temperature C 5 40 Permitted ambient humidity 80 Permitted o...

Страница 20: ...s lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ika cn POLAND IKA Poland Sp z o o Phone 48 22 201 99 79 eMail sales poland ika com USA IKA Works Inc Phone 1 910 452 7059 eMail...

Отзывы: