15
Calibrate the device:
In this menu option, you can calibrate the device�
Before starting the calibration, locate a reference measuring
temperature sensor at the centre of the cabinet�
Confirm to calibrate the device by pressing “Start/Stop” key
(I) and following screen appear:
Select 2-point calibration by pressing “Temp (+)” key (O)�
Following screen appear:
Confirm the selected calibration type by pressing “Start/
Stop” key (I)�
Switch to the first point calibration screen by pressing “Timer
(+)” key (K)�
Set the calibration temperature value by pressing “Temp (+)”
key (O) or “Temp (-)” key (P)�
Set the first point temperature (e�g� 100 ºC) by pressing
“Temp (+)” key (O) or “Temp (-)” key (P)�
Change the reference temperature value by pressing “Temp
(+)” key (O) or “Temp (-)” key (P)�
Example: 2-point calibration:
Switch to calibration screen by pressing “Temp (+)” key (O)
or “Temp (-)” key (P):
Confirm the setting by pressing “Start/Stop” key (I)� The de-
vice start to heat� The set temperature value changes to the
right side of the screen and actual temperature appears on
the left side of the screen:
Wait about 1–2 hours until temperature is stable (e�g� reach
about 98�8 ºC) and “Set“ LED (S) no longer flashes and the
following screen appear:
Input the calibration value from the reference measuring
device (e�g� 98�6 ºC) with “Temp (+)” key (O) or “Temp (-)”
key (P):
Confirm the value by pressing “Start/Stop” key (I)� Then, fol-
lowing screen appear:
Confirm the setting by pressing “Start/Stop” key (I) and
then the screen switches to reference temperature screen�
Confirm the setting by pressing “Start/Stop” key (I), the
screen switches to next point calibration�
The other points parameter can be set in the same way�
Back to edit calibration screen by pressing “Start/Stop” key
(I) again after confirming the settings�
Switch to the first point calibration screen by pressing “Timer
(+)” key (K)�
Press “Start/Stop” key (I) again to finish the first point cali-
bration�
The other point calibration can be performed in the same
way�
T mode:
T mode on:
the counter / timer starts to count when the
set temperature is reached�
T mode off:
the counter / timer starts to count when the
heating function is activated�