51
Fault rectification
Display
Description
What should be done?
What happens next?
No heating pulse in the inner
circuit.
No increase in temperature
within the inner water circuit.
Acknowledge the alarm.
Check the cooling water flow
Check the circulating pump.
Click on „Restart“ to repeat the
system test.
Click on „Next“ to skip the sys-
tem test.
No increase in temperature
within the outer water circuit.
Acknowledge the alarm.
Check the cooling water flow
Check the condenser and the
water inflow.
Click on „Restart“ to repeat the
system test.
Click on „Next“ to skip the sys-
tem test.
The function cannot be perfor-
med
The selected function cannot be
used because the temperature
of the cooling water has not
reached the required values.
Acknowledge the alarm.
Check the cooling water flow
Check the condenser and the
water inflow.
Check the circulating pump.
Click on „Restart“ to repeat the
system test.
Click on „Next“ to skip the sys-
tem test.
Drift criterion not achieved
This error message is generated
during a measurement of the
„Equalise“ or „Main Test“ status
if the drift conditions for the cur-
rent function of the C 6000 can-
not be met.
Acknowledge the error message
to revert to the „Wait“ message.
Switch to the „Servicing“ tab
and press the „Stirrer“ button to
check that the stirrer is working.
Perform a calibration of the C
6000.
Establish the specified ambient
conditions.
Содержание C 6000 global standards
Страница 1: ...IKA C 6000 global standards IKA C 6000 isoperibol 20000004212 Operating instructions EN C6000_092016...
Страница 2: ......
Страница 43: ...43 Sound Activation of the speakers...
Страница 61: ......
Страница 62: ......
Страница 63: ......