Error messages and troubleshooting table
Fault
Description
Causes
Effect
Solutions
Er 2
In remote operation (PC)
mode with activated
watchdog function in
mode 1: no communica-
tion between PC and RET
control-visc
• PC does not transmit any data
within the set watchdog time
• Connection to PC interrupted
Heating switched off
Motor switched off
• Change watchdog time
• Transmit data from PC within set
watchdog time (OUT_WDx@m)
• Check cable and plug
Er 3
Device internal tempera-
ture exceeds 80 °C
•
Heat accumulation between
heating plate and housing
•
Permitted ambient temperature
exceeded
Heating switched off
• Switch device off, leave to cool and
switch on again
• Change experiment
• Observe maximum permissible am-
bient temperature
Er 4
Difference between target
and actual speed exceeds
300 rpm
• Motor blocked or overloaded
Heating switched off
Motor switched off
• Reduce load torque or use smaller
magnetic rods
• Reduce target speed
Er 5
No temperature increase
in sensor at continuous
temperature difference
(selected time in menu)
• Sensor not in medium
• Volume of medium to be mea-
sured too large
• Heat conductivity of medium to
be measured too low
• Heat conductivity of the vessel is
too low
• In the case of indirect heating,
the overall heat conductivity re-
sistance is too large
Heating switched off
• Place the sensor in the medium
• Reduce the volume of the media
• Use a carrier fluid with better heat
conductivity properties
•
Replace the glass vessel with a
metal pot
• Increase the “Time-out” period
Er 6
Interruption in safety cir-
cuit
• Defective connection cable
Heating switched off
• Replace cable
Er 13
SAFE TEMP P adjustable
safety circuit defective
•
Target/current different of the
adjustable safety circuit for mini-
mum temperature monitoring
Heating switched off
•
After switching on, change the
SAFE TEMP to a different value; if
this solves the issue, the previous
value can be reset by switching the
device off and on again
Er 14
Short circuit in temperatu-
re sensor 1
• Short circuit in temperature sen-
sor plug
• Short circuit in the cable or tem-
perature sensor
Heating switched off
• Check the plug
• Replace the temperature sensor
Er 21
Fault during heating plate
safety test
• Safety relay does not open
Heating switched off
• Switch device off, leave to cool and
switch on again
Er 22
Fault during heating plate
safety test
•
S_CHECK cannot generate
H_S_TEMP
Heating switched off
• Switch device off, leave to cool and
switch on again
Er 24
The temperature of the
heating plate exceeds
SAFE TEMP H (Hotplate)
• SAFE TEMP H (Probe) has been
set to lower than the current tem-
perature of the heating plate
•
Disconnection of heating plate
control temperature sensor
Heating switched off
• Leave the heating plate to cool
• Set the SAFE TEMP H (Hotplate)
higher
Troubleshooting
The device will indicate that there is a fault by displaying an error
message “Er X” (see table) on the screen.
Heating will be switched off permanently (exception: WD - see
table).
The stirring function continues to operate at the pre-set speed if
there is a fault in the heating circuit (exception: WD - see table).
In order to reset the device after a fault, it may need to be discon-
nected from the power supply for several seconds. The period the
device is switched off may also need to be longer if the tempera-
tures need to be counterbalanced (residual heat display may no
longer be active; see “Residual heat display”).
The safety circuit test takes place within 20 seconds of switching
on the device; after the test, the other monitoring functions are
activated.
If a fault cannot be eliminated directly, you must perform a
RESET
(see “Settings” menu)
. If the faults still cannot be elimi-
nated, the appliance must be inspected by a technical service
technician.