10.2.8 Adjust the overflow prevention [cOP]
Only adjust the overflow prevention OP when the unit is installed.
If possible, carry out the adjustment when the tank is empty.
The tank may be partly filled.
►
Make sure that the overflow prevention OP is
not
covered by the
medium. Observe the minimum distance between the overflow
prevention OP and the level (→ Table 10-1).
►
Select [cOP].
►
Press [SET] and keep it pressed.
> [cOP] flashes for some seconds; then the continuous display indicates
that the adjustment is being made.
> If the adjustment is successful, [donE] is displayed.
►
Confirm with [Mode/Enter].
> If the adjustment is not successful, [FAIL] is displayed.
►
If necessary, lower the level or correct the position of the overflow
prevention [OP] and repeat the adjustment operation.
cOP
Minimum distance between the overflow prevention OP and the level during
adjustment:
Table
10-1
[cm]
[inch]
LT8022
2.0
0.8
LT8023
3.5
1.4
LT8024
5.0
2.0
The position of the overflow prevention OP can be determined by calling up
the parameter [OP]. Note the offset if necessary.
The current level is to be determined manually since the unit is not yet
ready for operation before the adjustment.
When the overflow prevention is activated ([OP] = [value]), an adjustment
[cOP] must be carried out each time:
• [MEdI] or [OP] were changed. In this case
is displayed.
• The installation position (height, orientation) was changed.
• The connection between the sensor and the tank ground (e.g. cable
length) was changed.