All maintenance procedures should be done by an experienced and qualified technician with
demonstrated knowledge and skills (electrical and mechanical) in the service of medical
equipment.
✓
This individual must have access to this manual and the proper tools.
✓
Lubrication of this device is
not
required.
12.1.
RECOMMENDED PERIODIC PERFORMANCE
CHECKS
Daily
Inspect all external cables, controls, and
the tabletop for wear and damage.
Damaged cables must be replaced
promptly. The
Aspect ISO-Drive
uses a
medical grade power cord which is not
user serviceable. Replacement must be
performed only by a qualified service
technician.
Weekly
•
Check battery operation by
disconnecting the AC power and
running the tabletop up and down.
•
Run the table top through its full
range of motions to help keep the
actuators from sticking or freezing up.
Quarterly
Battery must be charged every 3 months
only while in storage.
Semi-annually
Inspect carbon fiber tabletop.
12.2.
SERVICE & REPAIR STATEMENT
Only qualified personnel should perform repairs on this equipment. Please read this
entire document before performing any diagnostics or repairs. Some procedures listed
require this device to be energized while repairs are performed; please exercise
extreme caution while working with electrical components. Always exercise
appropriate lockout/tag out procedure while performing any diagnostics and service
on the table.
12.3.
REINITIALIZE ACTUATOR CONTROLLER
If it is necessary to reinitialize the actuator controller, indications described in section 13.
Troubleshooting, follow these procedures
To clear errors press and hold the Trendelenburg Up and Trendelenburg Down buttons
simultaneously for 15 seconds, release and wait for 15 seconds.
To reinitialize the actuators, press and hold the Elevate Up and Elevate Down buttons for
15 seconds, release and wait 15 seconds, then press the Trendelenburg Down button to
12.
MAINTENANCE, SERVICE & REPAIR