Operator Manual
Aspect ISO-Drive
L100-3265 Rev. A
29
Immunity Test
EN/IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
ESD
EN/IEC 61000-4-2
±8kV Contact
±2kV, ±4kV, ±8kV,
±15kV Air
±8kV Contact
±2kV, ±4kV, ±8kV,
±15kV Air
Floors should be wood,
concrete or ceramic tile. If
floors are synthetic, the
relative humidity should be
at least 30%.
EFT
EN/IEC 61000-4-4
±2kV at 100 kHz
repetition frequency
for AC Mains
±1kV at 100kHz
repetition frequency
for Signal I/O parts Port
±2kV Mains
±1kV I/Os
Main power quality should
be that of a typical
commercial or hospital
environment.
Surge
EN/IEC 61000-4-5
±0.5kV, ±1kV Line to
Line
±0.5kV, ±1kV, ±2kV
Line to Ground
±0.5 kV, ±1 kV Line
to Line
±0.5kV, ±1kV, ±2 kV
Line to Ground
Main power quality should
be that of a typical
commercial or hospital
environment.
Voltage
Dips/Dropout
EN/IEC 61000-4-11
0 %
U
T; 0.5 cycle At 0°,
45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % UT; 1 cycle
70 %
U
T; 25/30 cycles
for 50 Hz and 60Hz,
respectively
Single phase: at 0°
0 %
U
T; 250/300 cycle
for 50 Hz and 60 Hz,
respectively
Single phase: at 0°
0 %
U
T; 0.5 cycle At
0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 1 cycle
70 %
U
T; 25/30 cycles
for 50 Hz and 60Hz,
respectively
Single phase: at 0°
0 %
U
T; 250/300
cycle for 50 Hz and 60
Hz, respectively
Single phase: at 0°
Main power quality should
be that of a typical
commercial or hospital
environment. If the user of
the
Aspect ISR family of tables
require continued operation
during power mains
interruptions, it is
recommended that the
Aspect
ISR family of tables
be powered
from an uninterruptible
power supply or battery.
Power Frequency
IEC 61000-4-8
Magnetic Field
EN/IEC 61000-4-8
30A/m, 50Hz or 60Hz
30A/m, 50Hz or 60Hz
Power frequency magnetic
fields should be that of a
typical commercial or
hospital environment.
Guidance and Manufacturer’s Declaration- Immunity, All Equipment and Systems:
Table 2
The
Aspect ISO-Drive
is intended for use in the electromagnetic environment specified
below.
The customer or user of the
Aspect ISO-Drive
should ensure that it is used in such an
environment.