Operator Manual
Aspect ISO-Drive
L100-3265 Rev. A
23
13.
TROUBLESHOOTING
Note: The motion of the Tabletop is fully controlled by user interface with switches,
buttons. In the event of a loss of these motions it is expected that the Table Top will
remain stationary without any unwanted movement.
Problem/Symptom
Possible Cause
Remedy
1. Table controls
are not functioning.
1. No Power.
2. Battery is depleted.
3. Emergency Stop
Pushbutton engaged.
4. Control connection.
5. Electromagnetic
Interference.
6. Actuators uncalibrated.
1. Check electrical outlet.
2. Connect to AC Power.
3. Reset Emergency Stop Pushbutton.
4. Service Controller connections
according to Service Manual.
5. Refer to Section 5 of this manual.
6. Reinitialize Actuator Controller with
Handset Control. (Section 12.3)
2. Table does not
elevate.
1. No Power.
2. Battery is depleted.
3. Emergency Stop
Pushbutton engaged.
4. Hand Control Failure.
5. Control connection.
6. Actuator(s)
uncalibrated.
7. Actuator Failure.
1. Check electrical outlet.
2. Connect to AC Power
3. Reset Emergency Stop Pushbutton.
4. Replace Hand Control.
5. Service Control connections
according to Service Manual.
6. Reinitialize Actuator Controller with
Handheld Control. (Section 12.3)
7. Replace Elevate Actuator according
to Service Manual.
3. Trendelenburg
motion does not
operate.
1. No Power.
2. Battery is depleted.
3. Emergency Stop
Pushbutton engaged.
4. Hand Control Failure.
5. Control connection.
6. Actuator(s)
uncalibrated.
7. Actuator Failure.
1. Check electrical outlet.
2. Connect to AC Power.
3. Reset Emergency Stop Pushbutton.
4. Replace Hand Control.
5. Service Control connections
according to Service Manual.
6. Reinitialize Actuator Controller
with Handheld Control. (Section 12.3)
7. Replace Trendelenburg Actuator(s)
according to Service Manual.