Operator Manual
Aspect ISO-Drive
L100-3265 Rev. A
12
•
Check that the ground pin on the electrical cord plug is in good condition before
each time that it is plugged in.
•
When the table is not connected to AC power, the table will automatically switch to
backup battery mode. The table should only be used under backup battery power
temporarily for necessary function if the external power supply is lost.
•
It is recommended that AC power be applied for a minimum of 8 hours every day to
keep a proper charge on the batteries and achieve maximum battery life. Batteries
are constantly being charged during normal use when connected to AC power.
When the table is not in use and not connected to AC power, the Emergency
Stop Pushbutton must be fully engaged to prevent the battery from being
discharged and to safely power off the equipment. Storage of the table for
longer than 3 months requires the table to be connected to AC power with the
Emergency Stop Button released to charge the battery.
7.2.
Preparation for Patient Use
•
New Installation: This equipment will need to be properly cleaned before patient use
as it will inevitably come into contact with contaminants during shipping, unpacking,
storage, and installation.
•
After Installation: This equipment will need to be properly cleaned between uses with
patients as it will inevitably come into contact with contaminants during procedures.
Refer to section 11 of this manual for cleaning instructions and approved cleaning
substances.
7.3.
Patient Loading
•
Patients must be loaded from the side of table. There is a possibility of
tilting, instability and/or tipping if patient is loaded onto the pedestal end of
Table or the imaging end of the Table.