17
MNPG179
MOVE THE HANDPIECE TO SCAN THE TREATED AREA.
-
Laser treatments must be carried out under the close supervision of the
operator and with the patient fully conscious and able to provide
feedback on the effects of the machine. IACER Srl cannot be held
responsible for accidents due to a failure to observe this requirement.
-
USE OF THE CONTROLS, ADJUSTMENTS AND PROCEDURES OTHER THAN
THOSE SPECIFIED IN THIS OPERATING MANUAL CAN CAUSE EXPOSURE
TO HAZARDOUS RADIATION.
-
The operator is responsible for checking that the emission head remains
in contact with the affected area, in order to prevent emission of the
laser into areas other than those to be treated.
-
It is advisable to carry out treatment only when the machine is in perfect
mechanical working order and the laser specifications are suited to the
purpose (refer to the table of technical specifications).
-
The device is intended for use exclusively with the tip in direct contact
with the skin, in order to ensure the correct distance and dimensions of
the laser beam and compliance with the energy settings configured on
the display.
For this reason, keep the handpiece in contact with the
affected area during emission of the laser
. After activating the
handpiece by means of contact with the plates, avoid moving it towards
or directing it at other areas. NEVER DIRECT THE HANDPIECE AT PARTS
OF THE BODY THAT ARE SENSITIVE TO LASER BEAMS, LIKE THE EYES. DO
NOT LOOK INTO THE BEAM OF THE HANDPIECE OR THE ONE DIRECTED
OR REFLECTED DURING TREATMENT.
-
Never leave the device running unattended and always turn it off after
use.
-
In order to avoid contamination of the work environment and/or of the
people involved in its use, only use laser handpieces cleaned and
disinfected with care after the previous treatment.
-
It is strictly forbidden to use the device in the presence of flammable
anesthetic mixtures and in environments that are rich in oxygen. IACER
Srl cannot be held responsible for accidents due to a failure to observe
this requirement.
-
NEVER COVER THE AIR VENTS
: THIS COULD AFFECT THE SAFETY OF THE
MACHINE WHEN RUNNING. IACER SRL CANNOT BE HELD RESPONSIBLE FOR
ACCIDENTS DUE TO A FAILURE TO OBSERVE THIS REQUIREMENT.
-
The operator is urged to check the installation of the electrical system
for the equipment before turning on the power switch.