IACER Srl
5
MNPG179
Characteristic
Specification
Programmable treatment time
Between 1 second and 30min
Free memory
10
External dimensions (width x height x depth) 18x18x35cm
Weight of the body of the machine
2Kg
Distance between the laser emission point
and the skin
Around 38 mm
Spot area
Around 10mm
2
±3%
Operating
conditions
environmental
temperature
From +10°C to +40°C
relative humidity
From 15 to 93% without
condensation.
atmospheric pressure
500-1060 hPa
Conditions
for
storage/transport
with the device in
the
packaging
provided by the
company
environmental
temperature
From +5 to +40 °C
relative humidity
From 15 to 93% without
condensation.
atmospheric pressure
500-1060 hPa
The useful life of the device is 10 years.
Содержание I-TECH 10000
Страница 1: ...USER MANUAL MNPG179Rev 2of03 07 2019 Laser therapy ...
Страница 2: ......
Страница 4: ...IACER Srl ii MNPG179 ...
Страница 10: ...IACER Srl 6 MNPG179 Device and command description ...
Страница 11: ...IACER Srl 7 MNPG179 Labels ...
Страница 44: ......