In order to guarantee high performance (e.g. in terms of the supply of
power, the condition of the optical parts, certification of the IR Spot, the
state of the optical coupling device) and the absolute safety of the operator
and patient, the LA8000/LA10000 laser must be subjected to a functional
test a minimum of once every 12 months, and electrical safety must be
assessed in accordance with the latest edition of standard IEC62353 a
minimum of once every 24 months. The tests must be entrusted to qualified
technicians.
The LA8000/LA10000 laser has a software counter which alerts the need for
the above-mentioned tests, when 960 hours of effective use of the laser are
reached.
The operator is responsible for checking the correct operation of the buzzer
during treatment. The buzzer alerts anyone else in the vicinity that therapy
is in progress.
CLEANING
No particular detergents are required to clean the equipment. For safety
reasons, ALWAYS turn off the main switch at the back of the equipment and
unplug it before carrying out any maintenance or cleaning operations.
It is strictly recommended not to use diluents, detergents, acid solutions,
aggressive solutions or flammable liquids to clean the external surfaces of
the machine and its accessories. The use of these substances, together with
improper use of the accessories, can cause irreparable damage to the
equipment and to the electrodes and render the warranty null and void.
Do not use abrasive chemical solvents or detergents for cleaning the
handpiece and lens: check the head of the handpiece used for treatment for
any cracks that could let liquid in. Always allow the solvents of adhesives and
cleaning solvents and disinfectants to evaporate before using the laser
device, especially in the case of flammable solutions, to avoid the risk of
igniting the endogenous gases.
The handle of the tip can be cleaned with a neutral product. Be careful when
cleaning the laser output: do not use any products but remove any
impurities with a micro-fiber cloth.
After cleaning the external parts of the box, dry them with care before
restarting the equipment.
Do not, for any reason, disassemble the equipment for the purpose of
cleaning or inspection: there is no need to clean inside the I-TECH LA8000