IACER Srl
37
MNPG179
Guidance and manufacturer’
s declaration
–
ELECTROMAGNETIC
IMMUNITY
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used
in such an environment.
Immunity test
IEC
60601-1-2
Test level
Compliance
level
Electromagnetic
environment
–
guidance
Portable and mobile RF communications equipment should be used no
closer to any part of the device, including cables, unless the recommended
separation distances calculated from the equation applicable to the
frequency of the transmitter are respected.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3V
rns
from 150kHz
to 80MHz
6V
rms
from 150kHz
to 80MHz in
ISM band
3V
rns
from 150kHz
to 80MHz
6V
rms
from 150kHz
to 80MHz in
ISM band
=
=
P
V
d
1
5
,
3
P
V
d
=
1
12
for ISM band
Radiated RF
IEC 61000-4-3
3V/m
from 80MHz
to 2.7GHz
3V/m
[E
1
]V/m
P
E
d
=
1
12
𝑓𝑟𝑜𝑚 80𝑀𝐻𝑧 𝑡𝑜 800𝑀𝐻𝑧
P
E
d
=
1
7
𝑓𝑟𝑜𝑚 800𝑀𝐻𝑧 𝑡𝑜 2.7𝐺𝐻𝑧
Radiated RF to
RF wireless
communication
equipment
IEC 61000-4-3
3V/m
from 80MHz
to 6GHz
3V/m
[E
1
]V/m
P
E
d
=
1
6
𝑓𝑟𝑜𝑚 800𝑀𝐻𝑧 𝑡𝑜 6𝐺𝐻𝑧
where
𝑃
is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and
𝑑
is the recommended
separation distance in meters (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic site survey, could be less
than the compliance level in each frequency range. Interference may occur
near equipment marked with the following symbol:
Содержание I-TECH 10000
Страница 1: ...USER MANUAL MNPG179Rev 2of03 07 2019 Laser therapy ...
Страница 2: ......
Страница 4: ...IACER Srl ii MNPG179 ...
Страница 10: ...IACER Srl 6 MNPG179 Device and command description ...
Страница 11: ...IACER Srl 7 MNPG179 Labels ...
Страница 44: ......