The "Free" section on the I-TECH LA8000/10000 offers the possibility of
customizing 10 protocols - 5 for "Trigger Point" mode and 5 for "Scanning"
mode; all changes can be saved in the internal memory. This means that the
operator can find his/her favorite settings even after a long period of
downtime.
Program selection
Here are reported the instructions regarding selecting the appropriate
program related to the pathology:
1.
in the main menu, use the UP and DOWN keys to select
PATHOLOGIES and press ENTER.
2.
The PATHOLOGIEs window appears. Use the UP and DOWN keys to
scroll through the list of recommended protocols with the specific
parameters.
3.
After selecting the pathology, press the ENTER key. In this section
it is possible to press the BACK key to return to the main menu,
scroll through the protocols (and view the details) using the UP and
DOWN keys, and press the ENTER key to confirm the settings. Some
of the protocols are divided into two phases; the first is in "trigger
point" mode while the second is in "scanning" mode. The operator
is free to choose whether to proceed with the second phase or not;
the settings of the second phase appear at the end of the first one
and the operator is prompted to either continue or end the session.
4.
When the ENTER key is pressed, the I-TECH LA8000/10000 prompts
the operator to confirm and start treatment. Pressing the ENTER
key again starts activation of the laser and calibration of the work
parameters. During this phase, it is essential that the fiber optic of
the tip is connected to the relative socket on the panel, the
interlock is enabled (when applicable) and the actuator is not
pressed by anyone or anything. If these requirements are not met,
the software does not activate the laser and a message flashes on
the screen prompting the operator to connect the tip and check
that the actuator has not been pressed. Only when these
requirements are met does the I-TECH LA8000/10000 activate the
laser and start therapy.
5.
The I-TECH LA8000/10000 emits the laser only when the actuator
is pressed. To pause the instrument during therapy, simply release
the actuator: emission of the laser automatically stops and the
Содержание I-TECH 10000