Page 9
2. Safety Instructions
These safety instructions are required by law.
They promote workplace safety and accident
prevention.
2.1 Guidelines for Safe Opera-
tion
2.1.1 Scope
Comply with the accident prevention regula-
tion „DGUV Regulation 3“ to prevent injury to
yourself and others. Beyond that, national
regulations apply without restrictions.
2.1.2 Unit control
Do not perform any work which compromises
the safety of the unit. Obey all safety instruc-
tions and warnings present on the unit.
In case of a malfunction or electrical power
disruption, switch off the unit immediately and
prevent a restart. Repair malfunctions
promptly.
Restricted use.
IEC 60335-1 stipulates as follows:
This device may be used by children of eight
years of age and above as well as by persons
with reduced physical, sensory or mental ca-
pabilities or lack of experience and knowledge
so long as they are supervised or have been
instructed regarding the safe use of the de-
vice and understand the hazards that may re-
sult from it. Cleaning and user maintenance of
the unit must not be undertaken by children
without supervision.
2.1.3 Unit Operation
Health hazard due to the inhalation of
aerosols.
During operation of the HygroMatik high-
pressure nozzle system HPS, it is not permit-
ted to stay in the humidifier chamber.
Health hazard
The demineralized water required for the
operation of the high-pressure nozzle system
is not suitable for drinking.
Risk of material damage!
The unit may be damaged if switched on
repeatedly following a malfunction without
prior repair.
Rectify defects immediately!
Regularly check that all safety and monito-
ring devices are functioning normally. Do not
remove or disable safety devices.
2.1.4 Electrical
Risk of electrical shock!
Hazardous electrical voltage!
Any work on the electrical system to be per-
formed by certified expert staff (electricians
or expert personnel with comparable training)
only.
During maintenance or installation work, the
device must be disconnected from the power
supply and secured against being switched
on again. The absence of voltage must be
ensured by a measurement.
After electrical installation or repair work, test
all safety mechanisms (such as grounding
resistance).