Table 16. Recommended Separation Distances
Frequency of
Transmitter
150kHz to 80MHz
80MHz to 800MHz
800MHz to 2.5GHz
Equation
d = 1.2
√
P
d = 1.2
√
P
d = 2.3
√
P
Rated Maximum Output
Power of Transmitter in
Watts
Separation Distance
in Meters
Separation Distance
in Meters
Separation Distance
in Meters
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the separation
distance can be estimated using the equation in the corresponding column, where P is the
maximum output [power rating of the transmitter in watts (W)] according to the transmitter
manufacturer.
Note:
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and refl ection from structures, objects, and people.
Recommended Separation Distances between Portable and Mobile RF
Communications Equipment and the Smartsigns
®
Liteplus.
(IEC60601-1-2)
Table 17. Cables (IEC60601-1-2)
Cables and Sensors
Maximum
Length
Complies with
Pulse oximetry cable
10 ft. (3 m)
-RF emissions, CISPR 11, Class B/
Group 1
-Harmonic emissions, IEC 61000-3-2
-Voltage
fl
uctuations/
fl
icker emission,
IEC 61000-3-3
-Electrostatic discharge (ESD), IEC
61000-4-2
-Electric fast transient/burst,
IEC61000-4-4
-Surge, IE 61000-4-5
-Conducted RF IEC 61000-4-6
-Radiated RF, IEC 61000-4-3
Software download cable, RS-232
serial, 9 pin “D”
10 ft. (3 m)
Non-terminated cable, RS-232, 9
pin “D”
10 ft. (3 m)
Reusable SpO
2
sensor
3 ft (0.91 m)
Manuf
actur
er’
s Dec
lar
ation