38
When NIBP Smart In
fl
ation is set to OFF via Service Mode: in MANUAL mode the cuff
pressure is in
fl
ated until the in
fl
ation value selected by user, In AUTO/STAT Mode, the
cuff pressure is in
fl
ated until the in
fl
ation value selected by user in initial measurement
and, from the next measurement, the in
fl
ation value for Adult and Pediatric patients will be
last Systolic BP value + 50mmHg and the in
fl
ation value for Neonatal patients will be last
systolic BP value + 30mmHg.
When NIBP Smart In
fl
ation is set to ON via Service Mode: in MANUAL mode, the suitable
in
fl
ation value for the Adult and Pediatric patient is automatically calculated during the
in
fl
ation, and the in
fl
ation value for the Neonatal patient is initial in
fl
ation value selected by
the user. In AUTO/STAT mode, the suitable in
fl
ation value for the Adult and Pediatric
patient is automatically calculated during the in
fl
ation in initial measurement and, from the
next measurement, the in
fl
ation value will be last systolic BP value +50mmHg. Also, the
in
fl
ation value for the Neonatal patient is initial in
fl
ation value selected by user in initial
measurement and, from the next measurement, the in
fl
ation value for Neonatal patient will
be last systolic value +30mmHg.
SYS
DIA
MAP
%SpO
2
/min
°F
°C
M
mmHg
kPa
BP
1
SYS
DIA
MAP
%SpO
2
/min
mmHg
BP
Figure 17. Initial Infl ation
Pressure
Initiating MANUAL mode of NIBP operation
1. Press the NIBP start/stop button.
A single blood pressure measurement will be made. As soon as an NIBP measurement
begins, the SYS display shows the current cuff pressure.
Systolic, diastolic, and MAP values are presented when the measurement is completed.
The measurements remain in the numeric display for 2 minutes or until another NIBP
cycle begins.
SYS
DIA
MAP
mmHg
kPa
BP
SYS
DIA
MAP
%SpO
2
/min
°F
°C
M
mmHg
kPa
BP
1
SYS
DIA
MAP
%SpO
2
/min
°C
mmHg
SYS
Systolic
Diastolic
MAP
DIA
MAP
mmHg
Figure 18. Manual mode of NIBP operation
NIBP Monitoring