58
MAP high and low alarm limits
With the monitor in the normal monitoring mode:
1. Press the Alarm button until MAP high alarm limit is displayed.
2. Leave the limit unchanged or press the Up/Down (+/-) selection buttons as
needed to change the limit to another value.
3. Press the Alarm button once again until MAP low alarm limit is displayed.
4. Leave the limit unchanged or press the Up/Down (+/-) selection buttons as
needed to change the limit to another value.
SYS
DIA
MAP
%SpO
2
/min
°F
°C
M
mmHg
kPa
BP
1
SYS
DIA
MAP
%SpO
2
/min
Figure 34. MAP High Alarm Limit Setting
SYS
DIA
MAP
%SpO
2
/min
°F
°C
M
mmHg
kPa
BP
1
SYS
DIA
MAP
%SpO
2
/min
Figure 35. MAP Low Alarm Limit Setting
Alar
ms and Limits