81
17.4
Measurement Parameters
NIBP
Pulse Rate
Pulse Rate Range
Adult/Pediatric 40 BPM to 200 BPM
Neonatal 40 BPM to 240 BPM
Pulse Rate Accuracy
±2 BPM or ±2%, whichever is greater
NIBP (Non-Invasive Blood Pressure)
Technique
Oscillometric Measurement
Measurement modes
AUTO, MANUAL and STAT
AUTO Mode
Automatic NIBP measurements at intervals of 1, 2, 3, 4, 5, 10,
15, 30, 45, 60, 90, 120, and 240 minutes
MANUAL Mode
Single measurement initiated by NIBP Start/Stop switch
STAT Mode
Series of consecutive measurements for 5 minutes
NIBP pressure measurement range
Systolic pressure range
Adult: 60 mmHg to 250 mmHg (8 kPa to 33.3 kPa)
Pediatric: 60 mmHg to 250 mmHg (8 kPa to 33.3 kPa)
Neonatal: 40 mmHg to 120 mmHg (5.3 kPa to 16.0 kPa)
Diastolic pressure range
Adult: 40 mmHg to 200 mmHg (5.3 kPa to 26.7 kPa)
Pediatric: 40 mmHg to 200 mmHg (5.3 kPa to 26.7 kPa)
Neonatal: 20mmHg to 90 mmHg (2.7 kPa to 12.0 kPa)
Mean pressure range
Adult: 45 mmHg to 235 mmHg (6 kPa to 31.3 kPa)
Pediatric: 45 mmHg to 235 mmHg (6 kPa to 31.3 kPa)
Neonatal: 30 mmHg to 100 mmHg (4 kPa to 13.3 kPa)
Pressure Display Accuracy
Meets ANSI/AAMI SP10:2002+A1:2003
Cuff Pressure Range
0 to 300 mmHg (0 to 40 kPa)
Initial Cuff In
fl
ation
Adult: - 120, 140, 160 (default), 180, 200, 220, 240, 260 mmHg
(15.9, 18.6, 21.2 (default), 23.9, 26.6, 29.2, 31.9, 34.5 kPa)
Pediatric: - 120 (default), 130, 140, 150, 160, 170mmHg
(15.9 (default), 17.2, 18.6, 19.9, 21.2, 22.6 kPa)
Neonatal: - 80, 90 (default), 100, 110, 120, 130, 140 mmHg
(10.6, 11.9 (default), 13.3, 14.6, 15.9, 17.5, 18.6 kPa)
Overpressure protector
Adult/Pediatric 300 mmHg (N.C.), 330 mmHg (S.F.C.)
Neonatal 150 mmHg (N.C.), 165 mmHg (S.F.C.)
Standards
ANSI/AAMI SP10:2002+A1:2003, IEC60601-2-30:1999
EN1060-1:1995, EN1060-3:1997 and EN1060-4:2004.
Speci
fi ca
tions