3. Lift print head assembly and chassis from inside paper tray. Pull paper tray
forward clear of assembly.
10.26
Paper Tray Refitting
1. Lift print head assembly and chassis. Insert paper tray ensuring that the
paper tray is located within the base tray guide.
2. Carefully refit chassis screws and tighten to 60cNm, seal with M-Coat-D.
Refit Power Supply Unit as detailed in 7.16. Align microswitch as detailed
in 7.24
3. Refit print head support plate as per 7.11 step 5 and align print head as
detailed in 7.12.
10.27
Paper Width Microswitch Removal
1. Remove main PCB as detailed in section 7.8
Remove 2 securing nuts, washer and lift switch clear of assembly.
10.28
Paper Width Microswitch Refitting
1. Refitting is the reversal of removal, ensuring paper tray is fully open during
refitting procedure.
10.29
Front Panel Connector Assembly Removal
1. Remove event marker, toco and US/FECG connectors from the main board
and remove cable ties.
2. Lift front panel connector assembly clear.
46
Issue: 2 Draft
Rev: a
614365
Pin No
Toco
Ultrasound
1
Orange
Red
2
White
White
3
Violet
Yellow
4
Black
Violet
5
Yellow
Brown
6
-
Orange
7
-
Black