3.6
Twins Functional Testing - BD4002
Fit both US transducers and toco transducer and test as detailed in section 3.2.
Only one FECG transducer can be used at any one time and should be tested as detailed
in section 3.4.
When full width paper(210mm) is loaded, the unit will print two FHR grids one above
the other and one toco trace.
With the normal paper(150mm) fitted, the two FHR traces are overlaid on one grid, and
identified periodically.