11.1
Error Codes
Under certain fault conditions the BD4000 will print error codes relating to faults
within certain sections of the circuitry.
When an error is detected and is first switched on, a printout will automatically be
generated as shown below. The unit will then ‘shut down.’
Line
No
Description
Comment
1
PCB S’No, date of
manufacture
Factory preset
2
S
oft
w
are version
Factory preset
3
L
anguage 1 (English),
S
cale 20 beats/cm, grid
on,
C
hart 3cm/min,
M
ode (internal use only)
User selectable
4
Current Time
User selectable
5
Current Date
User selectable
The error codes on the bottom of the printout indicate the following;
Error Code
Possible Cause
0
No fault
1
Boot failure on DSP
2
Ram U10 - U11
3
Ram U10
4
Ram U11
5
Real Time Clock Battery
Issue: 2 DRAFT
53
Rev: a
614365