10.23
Paper Tray Open Microswitch Removal
1. Remove microswitch connector and power supply as detailed in 7.15.
2. Remove microswich and plate securing screws. Removing one screw at a time
will prevent the securing plate from dropping under the chassis. Lift
microswitch clear of assembly and clean chassis of sealant.
10.24
Paper Tray Open Microswitch Refitting
1. Place microswitch over securing plate and fit screws loosely.
2. With paper tray fully closed adjust microswitch position to register paper tray
closed.
3. Pull the left hand side of the paper tray gently, the unit may display “Paper
tray open” briefly as the paper tray is held under tension, this message must
disappear when released.
4. The switch must detect as soon as the left hand side catch is depressed and the
tray released.
5. Tighten microswitch securing screws to 35cNm.
10.25
Paper Tray Removal
1. Remove Power Supply PCB as per 7.15. Remove 6 chassis mounting screws
and disconnect stepper motor plug.
2. Remove print head support plate as detailed in 7.10 steps 1-2. (Cont’d)
Issue: 2 DRAFT
45
Rev: a
614365