10.35
Toco Transducer Alignment
Equipment required
a) Power Supply
b) Digital Volt Meter
c) Weight - 1 x 100gm
1. Remove case top as detailed in 10.31.
2. Set up power supply voltage to 5.00Vd.c. +/-0.05V.
3. Connect transducer to power supply and DVM as shown in table above.
4. Turn transducer so that strain gauge is uppermost.
48
Issue: 2 Draft
Rev: a
614365
View from rear of socket
Pin No
Connection
1
5V Supply
2
-
3
GND
4
+Ve O/P
5
-Ve O/p