ULTRASOUND EQUIPMENT USER MANUAL
9
Safety precautions
Careful
1. Should prevent the probe far away from the sharp, hard object, it is forbidden to collision probe
acoustic window, otherwise it will cause damage of the probe.
2. Probe for a long time can cause fever, during the suspension diagnosis, should press the "frozen"
button to make the probe to stop working.
3.
The probe’s waterproof grade is IPX7
,
and the part of the probe which is soaked in the water
can not surpass 5mm than the acoustic window.
4. Use unqualified coupling agent can make the probe acoustic window from corrosion and damage
should be used for medical ultrasonic coupling agent.
5. Only used in the probe head coupling agent, carefully clean after use.
6. Probe is forbidden to be heated. If the probe temperature is lower than the room temperature, it
can be heated naturally before use.
7. Do not insert or take down the probe in the electric situation.
8. In using process, it is strictly prohibited to cover the heat emission hole of the instrument.
9. Please do not open the instrument shell by oneself, in order to avoid possible electric shock risk.
Any instrument maintenance and upgrade must be authorized by the manufacturer, training and service
personnel.
10. Sundry, water cup, etc. are not permitted on the operation panel.
11. The instrument can only use the probe of manufacturer configuration; otherwise it will cause
damage of equipment.
12. In using process, if an exception occurs, please shut down immediately.
Attention
1. Power supply circuit of the instrument and the bearing capacity of the power supply
socket range
is 110V~240 V / 1.5 A.
If use extended patch panel, please choose a product which with the safety
certification.
2. When using should be far away from the strong electric field, strong magnetic field devices and
high voltage equipment.
3. If the instruments used in conjunction with other medical equipment, please consult manufacturer,
shall be carried out under the guidance of manufacturer.
4. Do not use the unspecified accessories (such as insurance and protect the earth wire, etc.), in order
to avoid adverse impact on performance and security in the device.
5. After shutdown the instrument every time, should wait for at least 2 ~ 3 minutes to start up again,
so as not to damage the instrument.
6. The screen is fragile, please do not hit, in order to avoid break.