ULTRASOUND EQUIPMENT USER MANUAL
40
Chapter 7 Fault Checking and Ruled Out
7.1 Fault Checking Need Attention
When Fault check in, please observe the following precautions:
Fault Checking Need Attention
Warning
According to this chapter, if the problem can not be solved successfully, please send back the
instrument to repair. Do not attempt to repair or remove instrument shell.
7.2 Fault checking and ruled out
If there is a problem when you are using, please read this chapter. You may be able to solve the
problem by yourself easily. Failure analysis and elimination of sample are shown in form as follows:
NO.
Fault phenomenon
Checking item
Ruled out methods
1
Open the instrument
power switch, display
is not bright, fan does
not turn.
1.
Whether a power outage?
2.
The power cord plugs loose?
3.
Insurance tube fuse?
1.
Energized before using.
2. Plug the power cord plug
tightly.
3. Replace the new insurance
tube.
2
The image on the
screen is not clear.
Total gain, the TGC, display
brightness and contrast
adjustment is appropriate or not?
Adjust the total gain, the
TGC, display brightness and
contrast.
3
Intermittent
interference or
snowflake
interference stripe to
appear on the screen.
1. If there is a strong high
frequency equipment or electric
and magnetic field interference
around the instrument?
2. The power cord and probe plug
and socket contacts are in good
condition?
3. The earthling rod connect with
ground is good or not?
1. To avoid the high
frequency or high voltage
equipment, magnetic field
around.
2. Plug the power cord and
the probe socket of the
instrument in right way.
3. Well grounded as required.