www.humancaregroup.com
25
Closing the lift - Roomer S
ENG
1. Make sure that all cables are connected as
they were before.
2. Make sure no cables are pinched.
3. Follow the steps 1-6 in the “opening the lift
chapter” in reverse.
Remember to attach the battery cable
before closing the lift!
Don´t forget the rubber feet!
Reseting service alarm – Roomer S
ENG
After the service is finished the service alarm
should be reset.
• Connect the lift to a computer thru “Data
and service application”. In the software,
click service and log in with your login
details. Click on the button “Reset service
alarm”.