www.humancaregroup.com
29
Changing the battery - Roomer S
ENG
The battery pack consists of two 12 V, 2.3 Ah,
lead type batteries. The lifetime is up to 5 years
depending on charging and usage.
Replacing the battery:
1. Follow the steps 1-6 in the “opening the lift
chapter”.
Note! Make sure to remove the battery
cable that connects to the circuit board
(not the battery cables that is placed on
the battery).
2. Unscrew the metal bracket from the chassis
by using a 2,5 mm Allen key.
Warning! Do not remove the battery
cables show in the picture below! They
should stay on until the screw is loosened
to avoid short circuit.
3. Now - disconnect the battery cables located
on the battery.
4. Lift up the battery pack from the lift.
5. When the batteries is removed from the lift,
disconnect all four cables.