![human care HeliQ Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/human-care/heliq/heliq_service-manual_2182734024.webp)
24
24
www.humancaregroup.com
Opening the lift - Roomer S
1. Activate the emergency stop.
2. Turn the lift upside down. Be careful with
the belt stop sensor.
3. Remove the rubber feet closest to the
hanger bar on Roomer S lift.
4. Untighten the two screws (TX10) holding the
covers together. The screws are located at
the bottom of the holes for the rubber feet.
5. Turn the lift around and remove the cover by
angle it a bit towards the opposite side of the
control panel until you hear a “click”.
6. Always disconnect the batteries when the
cover is opened.
ENG