88
www.humancaregroup.com
Changing the belt - Altair
ENG
The belt can be changed without removing the
belt outlet.
Follow the steps 1-5 in the “opening the lift
chapter”.
1. Re-activate the emergency stop.
2. Unscrew the Opto-switch for the belt
tension guard and turn it 90° in order to
be able to freely operate the lift without
weight.
3. Drag the belt out all the way from the belt
roll.
4. Create a big loop.
5. Use for example a spatula to create a loop
on the other side.
6. Press “up” button on the hand control to
turn the belt roll and grab the loop.