![Hubner C-FLEX Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/hubner/c-flex/c-flex_owners-manual_2180887062.webp)
62 | 82
OWNERS MANUAL | C-FLEX | D0579-B DECEMBER 2019
System Interfaces
The laser combiner has several system interfaces, through which the operator can control all integrated lasers
simultaneously. The key switch can be used to toggle the lasers ON and OFF as described in section 2.3. The
remote interlock, when removed will result in all lasers being disabled. The 15 V power supply delivers power to
all integrated lasers via internal power distribution and conversion circuitry.
Laser Interfaces
Cobolt 06-01 Series
Cobolt 06-01 Series lasers integrated into the combiner have an interface for status LEDs as well as modulation
inputs. There are two SMA connectors on each interface panel, one for digital and one for analog modulation
inputs to the Cobolt 06-MLDs and 06-DPLs.
Cobolt 06-01 laser interface on C-FLEX
The C-FLEX back panel
has LEDs to indicate each laser’s status. When power is supplied to the laser head,
regardless of key-switch state, the temperature control elements will be active to reach set point values.
The status of the laser operation is given via LED indicators:
ON
Green
The system is emitting or armed for emission.
ERROR
Red
An error has occurred.
Содержание C-FLEX
Страница 1: ...OWNERS MANUAL C FLEX D0579 B DECEMBER 2019 C FLEX Compact Flexible Laser Combiner...
Страница 2: ...2 82 OWNERS MANUAL C FLEX D0579 B DECEMBER 2019...
Страница 12: ...12 82 OWNERS MANUAL C FLEX D0579 B DECEMBER 2019 Label Placement...
Страница 71: ...71 82 OWNERS MANUAL C FLEX D0579 B DECEMBER 2019 Heat sink Laser Controller...
Страница 72: ...72 82 OWNERS MANUAL C FLEX D0579 B DECEMBER 2019 AOM Driver AOM Driver Heat Sink...