16.
Adjust the position of the camera or pin hole to the reference laser. Record
the reference beam’s exact
centroid position and do not move the camera until alignment process is complete for all lasers.
Beam profile camera control software displaying the centroid position of the measured beam.
17.
Make note of the centroid position of the reference beam. While comparing the reference laser will give
a good indication of the beam overlap, to be complete each laser position must be compared to all of
the others. To download the beam ‘Alignment Verification Calculation Tool’ go to
www.coboltlasers.com/downloads
NOTICE
–
Be sure to secure the pin hole or camera in place on the optical table before aligning
the remaining lasers. If the pin hole or camera moves the alignment will not be successful.
18.
Stop emission of the reference laser.
19.
Replace the aperture mounted pin hole and power meter.
20.
Start emission of the next laser to be aligned. It is advised to start with the laser nearest to the aperture
(position 1) and progressing away from the aperture when optimizing all lasers, including the reference
laser if necessary.
Содержание C-FLEX
Страница 1: ...OWNERS MANUAL C FLEX D0579 B DECEMBER 2019 C FLEX Compact Flexible Laser Combiner...
Страница 2: ...2 82 OWNERS MANUAL C FLEX D0579 B DECEMBER 2019...
Страница 12: ...12 82 OWNERS MANUAL C FLEX D0579 B DECEMBER 2019 Label Placement...
Страница 71: ...71 82 OWNERS MANUAL C FLEX D0579 B DECEMBER 2019 Heat sink Laser Controller...
Страница 72: ...72 82 OWNERS MANUAL C FLEX D0579 B DECEMBER 2019 AOM Driver AOM Driver Heat Sink...