![Hubner C-FLEX Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/hubner/c-flex/c-flex_owners-manual_2180887023.webp)
23 | 82
OWNERS MANUAL | C-FLEX | D0579-B DECEMBER 2019
Connecting the AOM driver cables
This section describes how to properly connect the driver, delivered with an integrated AOM, to the laser head
interface on the C-FLEX, as well as the input signals to the AOM Driver from the delivered DC power supplies.
Details on how to connect modulation signals will be
covered
in section 6 : Controlling Emission with Modulation
Mode Operation. For each AOM integrated in the C-FLEX there is an AOM driver. If no AOMs are integrated,
proceed to the next section.
NOTICE
–
A 5V DC supply voltage must be connected to the digital modulation input and 1 V DC
supply voltage must be connected to the analog modulation of the AOM driver or there will be no
transmission through the AOM and no laser emission at the C-FLEX aperture.
1.
For each AOM integrated in the C-FLEX there is a specific driver, they are labeled with the laser serial
number. Separate the drivers and cables by the laser they belong to and use only the corresponding
equipment for each laser.
2.
Connect the ‘AOM Driver Cable’s Sub
-D 15 connector to the Supply and Control connector on the AOM
driver.
NOTICE
–
Always connect the RF Input cable between the laser head and the AOM driver before
providing power to the AOM.
Содержание C-FLEX
Страница 1: ...OWNERS MANUAL C FLEX D0579 B DECEMBER 2019 C FLEX Compact Flexible Laser Combiner...
Страница 2: ...2 82 OWNERS MANUAL C FLEX D0579 B DECEMBER 2019...
Страница 12: ...12 82 OWNERS MANUAL C FLEX D0579 B DECEMBER 2019 Label Placement...
Страница 71: ...71 82 OWNERS MANUAL C FLEX D0579 B DECEMBER 2019 Heat sink Laser Controller...
Страница 72: ...72 82 OWNERS MANUAL C FLEX D0579 B DECEMBER 2019 AOM Driver AOM Driver Heat Sink...