PVCHECK
42
2. In case the instrument detects that the calibrated resistance is
higher that the measured resistance, the instrument gives out a
long sound and displays the message: “Calibration not OK”
3. In case the instrument detects a resistance higher than 5
Ω
at its
terminals, it gives out a long sound, zeroes the compensated
value and displays the message
“Calibration reset”.
6.6.
LIST OF DISPLAY MESSAGES
MESSAGE
DESCRIPTION
Function not available
The selected function/characteristic is not available
Data not saved
The instrument was not able to save the data
Wrong date
Set a correct system date
RADIO transmission error
The instrument does not communicate via RF with external units
SOLAR-02: Firmware incorrect
Inconsistent SOLAR-02 FW. Update firmware.
Firmware incorrect
Unsuitable instrument FW. Update firmware.
Error 4: contact service
Internal instrument error
Database full
The number of modules added to the internal DB is > 30
Module already added
The name of the module has already been added to the DB
Memory full
Instrument memory full upon pressing the GO key
Error: Vmpp >= Voc
Check module settings inside DB
Error: Impp >= Isc
Check module settings inside DB
Error: Vmpp * Impp >= Pmax
Check module settings inside DB
Error: Alpha too high
Check module settings inside DB
Error: Beta too high
Check module settings inside DB
Error: Gamma too high
Check module settings inside DB
Error: Toll too high
Check module settings inside DB
Wait data analysis
Download data from SOLAR-02 and wait for the result of PV efficiency
test
Download error
Contact service department
Error while saving
Problems with the access to memory area
Remote unit undetected
The instrument does not detect any remote unit SOLAR-02
Unable to perform analysis
Problems with the data downloaded from SOLAR-02. Check settings.
Data unavailable
Generic error. Repeat the test.