PVCHECK
31
Fig. 8: Connection for IVCK test with direct
irradiance measurement
o
Fig. 9: Connection for IVCK test with irradiance
measurement through SOLAR-02
CAPTION:
E: Green cable
C: Blue cable
P: Red cable
N: Black cable
1. PV module/string
2. Main system earthing
3. Earthed metal structure in the system
4. Reference cell for irradiance measurement
5. Temperature sensor (if required)
6. Remote unit SOLAR-02
14. In the initial screen of
IVCK
mode, the following values are
displayed in real time:
Module
type of module being tested
Vdc
value of output voltage from module/string
Irr
irradiance (from a direct measurement or SOLAR-02
connected via RF)
Tc
modulo temperature (in MAN or AUX mode) and the
relevant measuring mode, or “- - -” in AUTO mode
In case, the symbol “
” of RF connection with
SOLAR-02
CAUTION
Upon pressing the
GO/STOP
key, different error messages can be displayed by the
instrument (see § 6.6) and, therefore, the test cannot be started. Check and eliminate, if
possible, the problem causing the error message before going on with the test.
Содержание PVCHECK
Страница 1: ...Manual PVCHECK EAN 8033772797467...
Страница 2: ...PVCHECK 2 PVCHECK...
Страница 56: ...v 181002 2018 10 02 11 20...