PVCHECK
37
CAUTION
Upon pressing the
GO/STOP
key, different error messages can be displayed by the
instrument (see § 6.6) and, therefore, the test cannot be started. Check and eliminate, if
possible, the problem causing the error message before going on with the test.
4. Press
GO/STOP
to start the test. In case no error conditions occur,
the instrument displays the message “
Measuring…
” as shown in
the screen to the side.
5. When measurement is complete, the instrument provides the
Ri(+)min value, i.e. the minimum value of insulation resistance of
the PV module/string (or of other metal masses) being tested,
continuously measured during the whole duration of measurement.
If the result is higher than the minimum limit set, the instrument
displays the message “
Outcome:OK
”; otherwise, it displays the
message “
Outcome:NO
” as shown in the screen to the side.
6. Press the
SAVE
key to store the test result in the instrument’s
memory (see § 7.2) or the
ESC/MENU
key to exit the screen
without saving and go back to the main measuring screen.
6.4.4. Measuring insulation
– STRING mode
CAUTION
The maximum current which can be measured by the instrument is 10A. Before
carrying out the measures of Insulation in "STRING" mode, always ensure that the
instrument is connected to ONE STRING and not more strings in parallel to avoid
possible damage of it
1. Position the cursor onto
M
Ω
by using the arrow keys (
,
) and
confirm with
ENTER
. The display shows the screen to the side.
2. Press the
ENTER
key, activate “
Settings
” and possibly change
the desired parameters (see § 5.4). The following parameters are
shown on the display:
Ins. Test
selected test voltage (250, 500 or 1000VDC)
Ri min
minimum limit threshold for insulation measurement
Mode
measuring mode: STRING
Vtest
real applied test voltage
Rp
final value of the measurement obtained by the parallel
of Ri(+) e Ri(-) values which is compared with
the Ri min limit
by the instrument
key
access to the second page with the measured values
of VPN, VEP and VEN voltages
3. Connect the instrument to the PV module/string being tested and to the main earth node of the