5.
Remove
the
handles
and
front
panel
by
following
these
steps
and
referring
to
Figure
2-7.
a.
Remove
the
trim
strip
from
the
handles
.
b.
Remove
the
screws
attaching
each
handle
to
the
analyzer
.
c.
Pull
the
analyzer
toward
you
until
it
extends
about
two
inches
over
the
edge
of
the
table
top
.
d.
Grasp
and
pull
the
front
panel
with
two
hands:
one
on
the
top-middle
of
the
panel,
and
the
other
on
the
bottom-middle
of
the
panel.
Figure
2-7.
Removing
the
Handles
and
Front
P
anel
6.
Reconnect
the
frequency
counter
and
adapter(s)
to
the
RF
OUTPUT
connector
.
7.
T
o
obtain
a
counter
reading
of
500
MHz
62500
Hz
or
better
,
adjust
R1
on
the
A3
fractional-N/reference
assembly
(accessible
through
a
hole
in
shield
between
J1
and
J3).
8.
Reassemble
the
analyzer
.
Adjustments
2-11