![HP 87114B Скачать руководство пользователя страница 101](http://html.mh-extra.com/html/hp/87114b/87114b_service-manual_161524101.webp)
T
able
4-1.
HP
8711B/12B/13B/14B
Self-T
ests
(continued)
T
est
T
est
Name
Applicable
Error
Messages
and
Notes
22
DIN
Keyboard
Interface
Errors:
8042
DIN
Keyboard
test
F
AILED
Problem
with
DIN
Keyboard
P
ower
Supply!
Check
Fuse
.
T
ests
DIN
keyboard
control
processor
.
Replace
CPU
board
or
check
clock
Y53,
processor
U13.
23
Centronics
Interface
Errors:
Centronics
port
test
F
AILED
.
Strobe
bit
error
.
Printer
reset
bit
error
.
Printer
select
bit
error
.
T
est
writes
to
and
reads
back
three
printer
control
output
lines
.
Replace
CPU
board
or
check
U2,
U3,
U35.
Note:
T
ests
numbered:
8,
11,
and
14-17
are
not
run
during
power-up
.
Error
Occurs
During
DSP
Initialization
or
Calculating
Coecients
Clear
the
Nonvolatile
SRAM
Cycle
the
analyzer's
line
power
.
While
the
analyzer
is
going
through
its
self
tests
(Main
and
SIMM
DRAM
tests
or
shortly
afterwards),
press
the
4
PRESET
5
key
a
few
times
.
The
analyzer
will
pause
after
performing
the
CPU
main
self-tests
and
ask
if
you
want
to
zero
the
nonvolatile
SRAM.
(Note
the
caution
message!)
Press
4
BEGIN
5
to
do
so
.
Continue
if
the
error
persists
.
Measure
the
5
MHz
Reference
Signal
The
DSP
requires
this
signal
from
the
fractional-N/reference
assembly
for
proper
operation.
This
signal
is
routed
on
the
backplane
board.
Make
sure
that
the
fractional-N/reference
assembly
is
pushed
all
the
way
in
and
making
good
contact
with
the
backplane
.
Measure
the
5
MHz
signal
at
J6-C24
on
the
backplane
board
with
an
oscilloscope
.
If
the
signal
is
present,
remove
the
CPU
board
and
the
shield
over
the
DSP
chip
(on
the
right
side
of
the
board).
Make
sure
that
the
TMS
320C25
microprocessor
is
properly
seated
in
its
socket.
If
the
problem
persists
,
refer
to
\T
esting
the
DSP
Circuitry"
to
check
the
DSP
,
or
replace
the
CPU
board.
If
there
is
no
5
MHz
signal
at
the
backplane:
Check
the
reference
board.
If
an
external
reference
signal
is
used,
make
sure
it's
a
10
MHz
reference
with
a
level
of
at
least
05
dBm.
If
the
internal
10
MHz
reference
is
being
used,
check
the
output
at
A3J3
to
verify
that
the
crystal
oscillator
is
working.
If
the
10
MHz
signal
is
not
present,
replace
the
A3
fractional-N/reference
assembly
.
4-12
T
roubleshooting
and
Block
Diagrams
Содержание 87114B
Страница 7: ......
Страница 19: ...Absolute Power Accuracy broadband Dynamic Accuracy narrowband 1 4 Performance T ests ...
Страница 39: ......
Страница 73: ......
Страница 89: ......
Страница 115: ......
Страница 116: ......
Страница 118: ...Figure 5 1 The Service Key Menus 5 2 Service Related Menus ...
Страница 132: ......
Страница 162: ......
Страница 164: ......
Страница 170: ......