![HP 87114B Скачать руководство пользователя страница 41](http://html.mh-extra.com/html/hp/87114b/87114b_service-manual_161524041.webp)
Correction
Constants
(CCs)
A
djustments
Note
Please
read
the
next
paragraphs
carefully.
When
you
switch
ON
the
analyzer's
line
power
,
the
analyzer
copies
the
current
correction
constants
from
the
CPU
EPROM
into
a
RAM
buer
.
When
you
run
the
adjustment
tests
,
the
analyzer
generates
correction
constants
and
saves
them
to
the
RAM
buer
only
(they
are
erased
if
the
analyzer's
line
power
is
switched
OFF).
T
o
transfer
the
newly
generated
correction
constants
from
the
RAM
to
the
EPROM,
perform
the
\Storing
and
Recalling
Correction
Constants"
procedure
in
Chapter
8.
Y
ou
should
also
use
the
procedure
in
\Storing
and
Recalling
Correction
Constants"
to
make
a
copy
of
the
correction
constants
.
This
allows
you
to
quickly
restore
the
adjustment
data
if
you
replace
the
CPU
board
or
update
the
rmware
.
When
you
are
performing
the
adjustment
procedures
,
you
should
periodically
store
the
correction
constants
to
a
disk
le
.
This
allows
you
to
restore
data
quickly
in
case
you
need
to
switch
OFF
the
analyzer's
line
power
before
completing
the
adjustments
.
Caution
If
the
analyzer's
line
power
is
switched
OFF
at
any
time
during
the
adjustments
,
you
must
reload
the
previously
stored
correction
constants
before
proceeding.
F
ailure
to
do
so
will
result
in
loss
of
previously
performed
adjustment
data.
The
adjustments
will
have
to
be
performed
again.
2-2
Adjustments
Содержание 87114B
Страница 7: ......
Страница 19: ...Absolute Power Accuracy broadband Dynamic Accuracy narrowband 1 4 Performance T ests ...
Страница 39: ......
Страница 73: ......
Страница 89: ......
Страница 115: ......
Страница 116: ......
Страница 118: ...Figure 5 1 The Service Key Menus 5 2 Service Related Menus ...
Страница 132: ......
Страница 162: ......
Страница 164: ......
Страница 170: ......