If
This
T
est
F
ails
Repeat
the
test
after
verifying
that
all
connections
are
secure
.
There
is
no
corresponding
adjustment
for
this
test.
V
erify
that
the
A5
receiver
assembly's
detector
diodes
are
functioning
properly
.
See
\Troubleshooting
the
A5
Receiver
Assembly"
in
the
\Troubleshooting
and
Block
Diagrams"
chapter
for
more
information.
Directivity
(includes
source
match
and
test
port
match)
These
tests
require
an
HP
85032B
(standard
or
option
001)
calibration
kit
for
a
50
analyzer
or
an
HP
85036B
calibration
kit
for
a
75
analyzer
.
Note
The
standard
economy
(E)
versions
of
these
kits
do
not
provide
the
necessary
components
to
perform
this
procedure
.
Directivity
is
measured
by
presetting
the
analyzer
,
then
attaching
a
known-good
load
to
the
RF
output
and
performing
a
reection
measurement.
This
measures
the
default
calibration
of
the
analyzer
.
This
value
is
characteristically
30
dB
.
A
user
calibration
will
provide
40
dB
of
directivity
,
but
for
this
test,
we
are
testing
only
the
quality
of
the
default
calibration.
The
typical
specication
is
30
dB
.
The
program
prompts
you
through
all
portions
of
the
test
and
displays
the
results
on
the
screen
of
the
analyzer
.
Source
match
is
measured
at
the
same
time
as
directivity
.
This
test
can
be
performed
manually
by
taking
the
following
steps:
1.
Switch
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Meas
Cal
on
OFF
under
the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Meas
Cal
Options
portion
of
the
service
menu
to
disable
correction.
2.
P
erform
a
one-port
reection
calibration.
3.
View
the
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Source
Match
array
under
the
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
View
Array
Master
portion
of
the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Meas
Cal
Options
menu.
Raw
directivity
can
also
be
viewed
by
selecting
the
corresponding
softkey
.
Do
not
confuse
this
with
the
default
directivity
above;
raw
directivity
will
be
much
lower
.
Input
port
match
is
measured
by
performing
a
one-port
cal
at
the
end
of
a
type-N
cable
using
the
female
standards
contained
in
the
cal
kit.
F
or
this
reason,
the
economy
version
of
the
cal
kits
cannot
be
used
since
they
do
not
contain
female
standards
.
Once
the
user
cal
is
completed
the
cable
is
attached
to
the
RF
input
port
and
the
input
match
is
measured.
Please
note
that
when
the
female
open
is
called
for
during
this
user
calibration,
the
extender
pin
must
be
placed
over
the
center
conductor
of
the
cable
.
If
This
T
est
F
ails
Directivity
is
mostly
dependent
upon:
the
quality
of
adjustment
#107
the
quality
of
reference
load
used
in
the
adjustment
the
quality
the
load
used
in
the
test
If
directivity
should
fail,
repeat
adjustment
#107
with
a
known
good
load
(better
than
40
dB
of
return
loss).
Source
match
is
mostly
dependent
upon
hardware
in
the
A5
receiver
assembly
.
Any
failures
will
probably
require
the
replacement
of
the
A5
receiver
assembly
.
T
est
port
match
is
solely
dependent
upon
the
A5
receiver
assembly
.
If
either
of
these
two
tests
fail,
rst
inspect
the
two
type-N
connectors
for
damage
,
especially
on
75
units
.
If
the
test
port
match
fails
,
also
suspect
the
A5
receiver
assembly
diodes
.
1-22
P
erformance
T
ests
Содержание 87114B
Страница 7: ......
Страница 19: ...Absolute Power Accuracy broadband Dynamic Accuracy narrowband 1 4 Performance T ests ...
Страница 39: ......
Страница 73: ......
Страница 89: ......
Страница 115: ......
Страница 116: ......
Страница 118: ...Figure 5 1 The Service Key Menus 5 2 Service Related Menus ...
Страница 132: ......
Страница 162: ......
Страница 164: ......
Страница 170: ......