COMMAND SUMMARY
through 31.
or power on does not effect this value. Default is
13. It is defaulted only when memory is initialized.
l
DIAGnostics:INSTrument:PRINter:ADDRess
l
DIAGnostics:INSTrument:PRINter:ADDRess?
Sets the HP-IB address of the printer to use during some of the
calibration procedures when the swept signal generator assumes
HP-IB control. *RST and power on do not effect this command. The
default is 1. The default value is set at memory initialization only.
l
DIAGnostics:IORW
Performs a write to the I/O Device number specified in the first
<num> and sets it to the value in the second <num>.
l
DIAGnostics:IORW?
Reads from the specified I/O device number and returns the response
data.
l
DIAGnostics:OUTPut:FAULt?
Returns a string of 16, and that are equivalent to the fault
display.
Bit 0 = PEAK
Bit 1 TRACK
Bit 2 = RAMP
Bit 3 = SPAN
Bit 4 =
Bit 5 = ADC
Bit 6 EEROM
Bit 7 = PWRON
Bit 8 = CALCO
Bit 9 = PLLZERO
Bit 10 = PLLWAIT
Bit 11 = FNXFER
Bit 12 = CAL YO
Bit 13 = CAL MAN
Bit 14 = TMR CNFLCT
Bit 15 = SEARCH
Operating and Programming Reference S-31
Содержание 8360 B Series
Страница 231: ...HP IB Henu See Also CONNECTORS HP IB Getting Started Programming in Chapter I H 2 Operating and Programming Reference ...
Страница 363: ...NF N 2 w z P w D I oa 0 i B 6 7 On s5 n 20 a g 9 i T2 ...
Страница 406: ...2b Menu Maps Menu Maps 2b 1 ...
Страница 407: ...t t m ALC Bandwidth Select Module Sense Low High Auto AUTO Front Rear None ALC MENU ...
Страница 412: ...rMENU SELECT 1 Fault Fault Info1 Info2 Fault Info3 Clear Fault S E R V I C E M E N U 6 ...
Страница 469: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...